OASIS Command Referenceshutcom |
shutcom - shutter calibration utility
shutcom [<shutter driver name>]
The shutcom is used to send shutter calibration commands, one at a time, to the shutter. The shutter driver name is the name of the shutter driver as returned by the dp command. If the can is configured for only one shutter, the shutter driver name is optional. The following commands are available:
Command Function ccw nnnMove in CCW direction for nnn ticks cw nnnMove in CW direction for nnn ticks setSet the current position as home (closed) position, and update drv_parms openCall the shutter open routine with the current drv_parms closeCall the shutter closeroutine with the current drv_parms whereRead the current position of the paddles CRExit Shutcal Typically, shutcal is used to move the shutter to its closed position using the 4,5,6 commands, and then setting the home position with the h command. The O and C commands may be used to check the operation of the shutter.
TIP: The counts value displayed corresponds to a potentiometer value that ranges from 0-4095; the closed position is usually approximately 2000. If this is not the case, the shutter paddle may need to be rotated on its shaft. If the counts value does not change or the shutter does not respond, it is often due to a disconnected cable on the OASIS controller.
Calibrate Shutter0:
shutcom shutter0