# Calibration file for PRR-620 Spectroradiometer # Calibration Date: 04 Jun 1996 # File last modified: 19 Nov 1998 by Pete Strutton # # Format of this file was chosen to be consistent with Biospherical cal file # # First line has one entry for each address tag, specifying number of # channels for that address tag # # There follows one line of calibration data for each channel as follows: # bank chan type name a b c cmin cmax units # # bank: Bank number, i.e., the address tag for a given bank of channels # chan: Channel number # type: Type of channel, as follows: 0 = unused, 1 = down irrad, 2 = analog, # 3 = up radiance, 4 = up irrad, 5 = quadratic, 6 = Old Satlantic cals # 7 = New Satlantic cals (b is multiplier) or MCP # Once a count is converted to a voltage, via fixed algorithm, # the voltage is converted to given units as follows: # Type 0: not converted or reported # Types 1 - 4: value = (volts - c) / b (a is unused) # Type 5: value = a + (b * volts) + (c * volts^2) # Type 6: value = a * (volts - c) / b # Type 7: value = a * b * (volts - c) # name: Name of channel # a: Calibration "a" value, unused for types 1-4 # b: Calibration "b" value # c: Calibration "c" value, unused for type 6 # cmin: unused # cmax: unused # units: Units that value decodes into #12 16 14 8 # Surface instrument, Serial 9664, Calibration date 4jun96 0 1 1 Ed412 0.0 -0.034257 -0.000370 0 10 uwatts/cm2/nm 0 2 1 Ed443 0.0 -0.033874 -0.000977 0 10 uwatts/cm2/nm 0 3 1 Ed490 0.0 -0.034658 -0.000318 0 10 uwatts/cm2/nm 0 4 1 Ed510 0.0 -0.034451 -0.000225 0 10 uwatts/cm2/nm 0 5 1 Ed555 0.0 -0.034199 -0.000238 0 10 uwatts/cm2/nm 0 6 1 Ed656 0.0 -0.032990 0.000045 0 10 uwatts/cm2/nm 0 7 2 PAR 0.0 -14.61530 -0.000178 0 10 uEinsteins/cm2/sec 0 8 2 Gnd 0.0 1.0 -0.000249 0 10 volts 0 9 2 Diode 0.0 0.1 0.0 0 10 degC 0 10 2 DGnd 0.0 1.0 0.0 0 10 volts 0 11 2 Pitch 0.0 0.0410 2.6786 0 10 degrees 0 12 2 Roll 0.0 0.0416 2.6549 0 10 degrees