# Calibration file for Satlantic Stor-dat sn 18 w OCR-100 sn 151 # Serial Number 018 # Calibration data from report dated 11 Sep 1197 # File created 27 Jul 2000 by Brian Schlining # First row: number of samples to skip, number of samples to convert # Following rows: Sensor, calibration factor, immersion factor, offset 0 10 Lu412 8.39E-05 1.748 32769 Lu443 8.89E-05 1.742 32767 Lu490 8.48E-05 1.736 32766 Lu510 8.54E-05 1.733 32768 Lu555 7.86E-05 1.729 32781 Lu670 3.13E-05 1.721 32771 Lu683 3.44E-05 1.720 32770 Ed490 1.02E-02 1.000 32771 Unused 1.000 1.000 0 Temp 1.000 -0.05427 1561.08 # Note for Temp calibration: the 1561.08 is derived from the given # calibration of 62.7628 as follows: # Satlantic equation: T = a0 + (a1 * cnt), a0 = 84.72, a1 = -0.05427 # Our equation: T = b0 * b1 * (cnt - b2) # So we set b0 = 1; b1 = a1 = -0.05427; b2 = -a0/a1 = 1561.08